1 Calibration Procedures
64 Keysight U3606B Service Guide
Calibration Process
The test equipment recommended for the calibration process and procedure is
listed in “Recommended Test Equipment” on page 20.
The following procedure is the recommended method to complete a full
instrument calibration.
1 Prior to performing the verification tests, see the “Test Considerations” on
page 25.
2 Perform the verification tests to characterize the instrument; see the
“Performance Verification Tests Overview” on page 38.
3 Unsecure the instrument for calibration; see “Unsecuring the Instrument for
Calibration” on page 66.
4 Prior to performing the adjustments, see the Adjustment Considerations” on
page 73.
5 Perform the adjustment procedure; see Adjustments Procedures” on page 75.
6 Secure the instrument against unauthorized calibration; see “Finishing the
adjustments” on page 99. Ensure that the instrument has quit the adjustment
mode and is turned off.
7 Record the new security code and calibration count in the instrument's
maintenance records.
NOTE
Ensure that you quit the adjustment mode before powering off the U3606B.