Calibration Procedures 1
Keysight U3606B Service Guide 73
Adjustment Considerations
1 Allow the instrument to warm up and stabilize for 3 minutes before performing
the adjustments.
2 Consider the thermal effects as you are connecting the test leads to the
calibrator and multimeter. It is recommended to wait for 1 minute before you
begin the calibration after connecting the test leads.
Never turn off the instrument during an adjustment. This may delete the
calibration memory for the present function.