Calibration Procedures 1
Keysight U3606B Service Guide 99
Finishing the adjustments
1 Remove all shorting plugs and connections from the instrument.
2 Reset the calibration message (“Calibration Message” on page 101).
3 Record the new calibration count (“Calibration Count” on page 100).
4 Press [Shift] > [Exit] or [Shift] + [] to exit the adjustment mode.
5 The instrument will be secured and returns to normal operation.