1 Calibration Procedures
98 Keysight U3606B Service Guide
For adjustment item “LOAd1”, “LOAd2”, “LOAd3”, and “LOAd4”, connect an
additional fixed resistor across the rear panel + and output terminals
according to the fixed resistors value in Table 1-33. Leave the connections
from the output terminals (+ and –) to the input terminals (V and LO) intact.
5 Verify the constant voltage adjustments using the “CV programming and
readback accuracy” on page 49.
NOTE
Do not remove the short bar between the rear panel sense terminals (+S and –S)
are the rear panel output (+ and –) terminals. See Chapter 3 of the U3606B
User’s Guide, “Remote sensing connections,” starting on page 116 for more
information on how to connect the load leads.
+S + + S
Fixed resistor