Calibration Procedures 1
Keysight U3606B Service Guide 41
Table 1-2 Zero offset verification test
Input Function Range Error from nominal 1-year
Short
DC voltage
19.9999 mV ±0.008 mV
100.000 mV ±0.008 mV
1.00000 V ±0.00005 V
10.0000 V ±0.0005 V
100.000 V ±0.005 V
1000.00 V ±0.05 V
2-wire resistance
100 Ω ±0.008 Ω
1000 Ω ±0.05 Ω
10 kΩ ±0.0005 kΩ
100 kΩ ±0.005 kΩ
1 MΩ ±0.00005 MΩ
10 MΩ ±0.0005 MΩ
100 MΩ ±0.005 MΩ
4-wire resistance
0.1 Ω ±0.00005 Ω
1 Ω ±0.0003 Ω
10 Ω ±0.003 Ω
100 Ω ±0.03 Ω
1000 Ω ±0.3 Ω
CAUTION
Zero offset calibration using a multifunction calibrator is NOT recommended.
The calibrator and cabling offset can be large and unstable causing poor offset
calibration of the Keysight U3606B or any multimeter.