1 Calibration Procedures
90 Keysight U3606B Service Guide
CC output adjustment procedure — front output terminals
Review the “Test Considerations” on page 25 and “Input and Output Connections”
on page 26 sections before beginning this procedure.
1 Turn off the instrument and connect the front panel (red) and (black)
output terminals to the I (red) and LO (black) input terminals.
2 Turn on the instrument. Press Current to enter the constant current output
calibration.
3 The display shows the uncalibrated value and the reference value of the
adjustment item.
4 Configure each adjustment item shown in Table 1-32.
5 Use or to select the adjustment item.
NOTE
Always complete tests in the same order as shown in Table 1-32.