31 32
Press and hold the Power button for 5 seconds or
longer and display the Setup screen. Then press
Up/Down button to select "Initialization" and press
START.
Select Yes using Up/Down buttons, and press START.
5-8. Initialization (Restoration to factory setting)
Time and date, past measurement data, and Bluetooth registration are deleted and the
product is reset to the factory shipment status. The calibration data are reset to the
factory shipment status.
The time and date cannot be set with this product alone. To set the time and date, an
application-installed smartphone/tablet or software-installed computer must be
connected.
The time and date can then be set automatically.
6-2. Maintenance of measurement chamber
If the reagent is spilled in the measurement chamber, clean the measurement
chamber.
In addition, clean the measurement chamber approximately once every six
months.
Perform self-check as needed. For self-check, refer to "5-3 Self-Check" (
P27).
Turn the power o.
Open the measurement chamber cover.
Thoroughly wipe the bottom and the side of the measurement chamber with
the provided cleaning brush soaked with ethanol.
Close the measurement chamber cover.
6 Maintenance
6-1. Maintenance of the main body
If the product becomes contaminated, wipe with a dry cloth or soft paper such as
tissue paper.
If the product becomes heavily contaminated, wipe with a soft cloth soaked with
rubbing alcohol or diluted mild detergent solution.
• Do not pour ethanol into the measurement chamber.
• Do not apply ethanol to any part except for the measurement chamber.
• Do not use the product until it dries.
• Do not spill liquids, reagents, organic solvents, etc. over the product.
If any liquid is spilled on the product, wipe the liquid o immediately, remove the
batteries, and allow the product to dry for 24 hours or longer indoors.
Initialization
No
Yes
Initialization
No
Yes
Caution
Caution