Kikkoman Lumitester PD-30 - User Manual

Kikkoman Lumitester PD-30
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Table of Contents
1. Read This First
1.1 Instructions for Use
1.2 Software Requirements
1.3 Description of Functions
2. How to install the Software
2.1 Before Installation
2.2 Installation Procedure
  2.2.1 Install the USB Driver
  2.2.2 Install the Control Software
3. How to operate the Software
3.1 Start-up and Exit the Software
3.2 Basic Operation
  3.2.1 Language Selection
  3.2.2 Date/Time Set
  3.2.3 User Name Input
3.3 PLAN Measurement
  3.3.1 Level Set
  3.3.2 PLAN Set
  3.3.3 PLAN Measurement Procedure
3.4 Memory Data Management
  3.4.1 Loading Memory Data
  3.4.2 Saving Memory Data in File
  3.4.3 Printing Memory Data
  3.4.4 Memory Data Clear
  3.4.5 Summarize Memory Data
3.5 AX+B Measurement
  3.5.1 AX+B Measurement Setting
  3.5.2 AX+B Measurement Procedure
3.6 Product Information Screen
3.7 Error Message
3.8 Troubleshooting
4. How to uninstall the Software
1
1
1
2
2
2
3
3
7
10
10
12
12
13
13
14
14
16
19
20
20
21
23
23
23
26
27
28
29
30
31
32
ttttttttttttttttttttttttttttttttttttttttttttttt
ttttttttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttttt
ttttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttttttttt
ttttttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttttttttttt
ttttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttttttt
ttttttttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttt
ttttttttttttttttttttttttttttttttttt
ttttttttttttttttttttttttttttttt
ttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttt
ttttttttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttttttttttttt
tttttttttttttttttttttttttttttttttttttttttttt
ttttttttttttttttttttttttttttttttt
Thank you very much for purchasing the Lumitester PD-30.
All of this Instruction Manual must be read before
operation of this product for safe and proper use. This
Instruction Manual should be kept for future reference.
Instruction Manual
Lumitester PD-30
Control Software

Questions and Answers

  • A
    Ann CobbAug 16, 2025
    What to do if the Kikkoman Lumitester PD-30 shows a water hazard?
    • Z
      Zachary RoseAug 16, 2025
      If the surface of the instrument was exposed to water or water has entered the measurement chamber: * Turn the power off immediately. * Wipe away any water from the instrument, or remove the reagent and wipe off water with the cleaning brush if water entered the measurement chamber. * Detach the battery cover, remove the batteries, and dry the instrument with the operation panel turned upward and the measurement chamber cover opened. * Allow the instrument to stand at room temperature for approximately 24 hours.
  • G
    gomezanthonyAug 19, 2025
    Why does Kikkoman Test Equipment power automatically turn off?
    • J
      Joshua SantiagoSep 12, 2025
      The Kikkoman Test Equipment might automatically turn off because the batteries are exhausted. In this case, insert new batteries. The device is also designed to automatically turn off after 10 minutes of inactivity, which is normal operation.
  • R
    ryanmartinezAug 22, 2025
    What to do if the Kikkoman Test Equipment power cannot be turned on?
    • E
      Emily HumphreyAug 22, 2025
      If the Kikkoman Test Equipment fails to power on, the batteries may not be inserted or may be exhausted. Try inserting new batteries.
  • J
    Jeffrey NashAug 25, 2025
    What to do if the Kikkoman Lumitester PD-30 displays E021 to E029 Measurement error?
    • K
      Karen SanchezAug 25, 2025
      The E021 to E029 measurement error on your Kikkoman Test Equipment may occur due to large temperature variations, direct sunlight exposure, or a loose measurement chamber cover. To resolve this: * Turn the power off. * Allow the instrument to stand for at least 30 minutes at room temperature if it was moved from a hot or cold location. * Move the instrument away from direct sunlight. * Ensure the measurement chamber cover is securely closed.
  • V
    Veronica ObrienAug 27, 2025
    What to do if the Kikkoman Lumitester PD-30 displays E031 to E039 Calibration error?
    • A
      Amy LeAug 27, 2025
      The E031 to E039 calibration error on your Kikkoman Test Equipment may occur due to large temperature variations, direct sunlight exposure, or a loose chamber cover. To resolve this: * Turn the power off. * Allow the instrument to stand for at least 30 minutes at room temperature if it was moved from a hot or cold location. * Move the instrument away from direct sunlight. * Ensure the measurement chamber cover is securely closed.
  • R
    Ricky EdwardsAug 30, 2025
    How to resolve E040 to 049 Temperature error on Kikkoman Lumitester PD-30 Test Equipment?
    • Y
      ydennisAug 30, 2025
      The E040 to E049 temperature error on your Kikkoman Test Equipment can occur if the temperature is out of range or due to large temperature variations. To fix this: * Turn the power off. * Move the instrument to a location with a temperature between +5 to +40°C. If temperature compensation is enabled, ensure the temperature is between +10 to +40°C. * If the instrument was moved from a hot or cold location, allow it to stand at room temperature for at least 30 minutes before use.
  • L
    Lauren MoodySep 2, 2025
    What should I do if the Kikkoman Test Equipment power cannot be turned off?
    • E
      Emily BlackburnSep 2, 2025
      If the Kikkoman Test Equipment is stuck on, try reinserting the batteries. Also, if a USB cable is connected, exit the control software and disconnect the USB cable.
  • J
    Julie RogersSep 5, 2025
    How to fix E051-059 Instrument error on Kikkoman Test Equipment?
    • S
      Sarah HernandezSep 5, 2025
      The E051-059 instrument error on your Kikkoman Test Equipment might be caused by malfunctions of electric parts. To solve this, turn the power off once, and then turn it on again.
  • S
    susan09Sep 7, 2025
    What to do if measured values seem lower on Kikkoman Lumitester PD-30 Test Equipment?
    • J
      Jody BurnsSep 7, 2025
      If the measured values appear to be lower than expected on your Kikkoman Test Equipment, the measurement chamber may be contaminated. Perform the "6.2 Maintenance of measurement chamber" procedure to clean it.
  • J
    James ReedSep 10, 2025
    How to resolve E011 to 019 Memory error on Kikkoman Lumitester PD-30 Test Equipment?
    • C
      Christopher HuntSep 10, 2025
      The E011 to E019 memory error on your Kikkoman Test Equipment might be caused by removing the batteries while memory data is being written or read. Try turning the power off and then on again. If the error persists, perform "5.2.2 Memory Data Clear".

Summary

Read This First

Instructions for Use

Software usage guidelines, compatibility, and connection precautions for Lumitester PD-30.

Software Requirements

Specifies PC hardware and operating system requirements for the Lumitester control software.

Description of Functions

Outlines the software's capabilities, including data retrieval and measurement condition setting.

Software Installation Guide

Pre-installation Checks

Details prerequisites and steps required before installing the software and USB driver.

Installation Procedure

Step-by-step guide for installing the Lumitester PD-30 control software.

Install USB Driver

Instructions for installing the necessary USB driver for PC connection.

Install Control Software

Steps to install the Lumitester PD-30 Control Software application.

Software Operation Guide

Start-up and Exit Software

Guide on how to launch and properly close the Lumitester PD-30 Control Software.

Basic Software Operations

General instructions for operating the software and its fundamental functions.

Language Selection

How to select and set the display language for the software and the PD-30.

Date and Time Settings

Instructions for setting the date and time on the PD-30 unit via the control software.

User Name Input

Procedure for inputting and managing user names within the software interface.

PLAN Measurement Setup

Guide to setting up PLAN measurements, including defining levels and object names.

Level Set Configuration

How to configure level sets for PLAN measurement, including rank judgment criteria.

PLAN Creation

Steps to create and define a PLAN for series of measurements.

PLAN Measurement Procedure

Detailed steps for executing a PLAN measurement using the software.

Memory Data Management

Overview of managing measurement data, including loading, saving, and clearing.

Loading Memory Data

How to load measurement data from the PD-30 into the control software.

Saving Memory Data to File

Instructions for saving measurement data in CSV or JPEG file formats.

Printing Memory Data

Procedure for printing the loaded memory data from the software.

Memory Data Clear

Steps to delete all stored memory data from the PD-30 unit.

Summarizing Memory Data

How to extract and summarize data based on specified conditions for analysis.

AX+B Measurement

Introduction to AX+B measurement, its purpose, and conversion formula.

AX+B Measurement Settings

How to configure settings for AX+B measurement, including units and coefficients.

AX+B Measurement Procedure

Detailed steps for performing AX+B measurements using the software.

Product Information Screen

Error Message Guide

Troubleshooting Common Issues

How to Uninstall Software

Kikkoman Lumitester PD-30 Specifications

General IconGeneral
BrandKikkoman
ModelLumitester PD-30
CategoryTest Equipment
LanguageEnglish