3 4
1
Names and Functions
1-2. Display Unit
1-1. Product
Display Unit
Operation Unit
Power Button
Stand
Measurement
chamber
USB port
Measurement
chamber cover
Battery cover
Communication indicator
Connection indicator
Bluetooth ON indicator
Reagent warning or measurement
chamber cover warning
Battery indicator
Measured value
Memory number
Temperature compensation
indicator
Measurement error indicator for temperature compensation enabled
Target
Bluetooth
USB
Display
BLE
USB
#0123
E
456789
BLE
USB
Battery indicator
Full
High
Medium
Low battery
Display
Display Instruction
Insert the reagent or close the measurement
chamber cover.
Remove the reagent.
Display Unit
Operation Unit
Power Button
Stand
Measurement
chamber
USB port
Measurement
chamber cover
Battery cover
Communication
indicator
Connection indicator
Bluetooth ON indicator
Battery indicator
Measured value
Memory number
Temperature
compensation indicator
Measurement error
indicator for
temperature
compensation
enabled
Reagent warning or
measurement
chamber cover
warning
Indicates communication with smartphone, tablet, and/or PC.
Indicates connection with smartphone,
tablet, and/or PC.
Indicates that the Bluetooth ON/OFF switch of the product is ON.
Displays the remaining battery level.
The measured value is displayed up to 6 digits with right
justication.
Shows the measured data number with # + 4 digits.
Indicates that the temperature compensation setting is ON.
If the temperature compensation is ON, the error code "E047" or
"E048" is displayed at the temperature 40°C or higher or 10°C or
lower, respectively, during measurement countdown, and the
measurement result without temperature compensation is
displayed with the error "E".
Indicates that you must close the measurement chamber cover,
or remove or insert the reagent.
Panel that displays the results of measurement, the
mode number, etc.
Refer to "1-2. Display Unit" (
P4).
Panel for measurements or setting operation.
Refer to "1-3. Operation button" (
P5).
Turns ON/OFF the power.
Displays the setup menu.
Used to set the product in an upright position.
Chamber where reagent is inserted.
Cover for the USB connector. Connector that is used
for PC connection.
Cover for measurement chamber.
Cover used to insert and remove batteries and switch
Bluetooth ON/OFF.

Related product manuals