NOTICE!
Danger of damage to the melt water container!
The units in the table below are equipped with a melt
water container on their underside, which can be dam-
aged by incorrect transport:
Place unit on pallet and transport it.
Do not lift unit with forklift or pallet truck unless it is
on a pallet.
LABO 520 PRO-ACTIVE LABO 520 ULTIMATE
LABO 720 PRO-ACTIVE LABO 720 ULTIMATE
LABO 520 CR PRO-ACTIVE LABO 520 ULTIMATE
LABO 720 CR PRO-ACTIVE LABO 720 ULTIMATE
LABO 720 CHROMAT PRO-ACTIVE LABO 720 CHROMAT ULTIMATE
FROSTER LABO 530 PRO-ACTIVE FROSTER LABO 530 ULTIMATE
FROSTER LABO 730 PRO-ACTIVE FROSTER LABO 730 ULTIMATE
FROSTER LABO 530 CR PRO-ACTIVE FROSTER LABO 530 CR ULTIMATE
FROSTER LABO 730 CR PRO-ACTIVE FROSTER LABO 730 CR ULTIMATE
Protective equipment:
n
Safety shoes
n
Protective gloves
Decommission the unit as shown in section
Ä
‘Final decommis-
sioning of the unit’ on page 35.
1. Transport the unit to the new location.
Waiting time before recommissioning:
Once the unit has been transported, set up the
unit and wait one hour before recommissioning,
so that the refrigerant can distribute itself evenly
throughout the refrigerant circuit again.
2.
Recommission the unit as shown in section
Ä
‘Putting the
unit back into operation’ on page 35.
Transporting the unit to a new loca-
tion
Transport and decommissioning
Unit transport
23.05.2023Refrigerator/freezer LABO series34