Kirsch LABO Series - Temperature Warning Limits; Function of the Temperature Warning Limits

Kirsch LABO Series
81 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Consequences of modified target temperature
As soon as the target temperature is reached, the tem-
perature warning limits must be adjusted so that the
target temperature is above or below the temperature
warning limits. Otherwise the temperature alarm will be
triggered
Ä
Chapter 6.2.2.2 ‘Displaying and changing
the temperature warning limits’ on page 42.
The set target temperature is saved automatically. The
target temperature is retained after a power failure and
after the unit has been switched off.
7. When the unit has reached its target temperature, stock the
unit.
6.2.2 Temperature warning limits
6.2.2.1 Function of the temperature warning limits
The temperature warning limits define how much deviation the unit
will tolerate between the actual temperature and the target temper-
ature. The temperature warning limits are unit-specific. They can
be adapted to the requirements of the chilled goods.
The values for the temperature warning limits are set at the factory
and comply with the valid DIN standards for the unit.
Recommended temperature warning limits
The temperature warning limits must not be the same
as the target temperature.
Set the temperature limits as follows:
Upper temperature warning limit: at least 3 °C
higher than the target temperature
Lower temperature warning limit: at least 3 °C lower
than the target temperature
Recommended temperature warning limits for
FROSTER
To ensure that the unit functions properly, the tempera-
ture warning limits must be set for FROSTER models
as shown in table
Ä
‘Temperature warning limits’
on page 42.
Commissioning
Programming the unit > Temperature warning limits
23.05.2023 Refrigerator/freezer LABO series 41