13.2 Technical data
LABO 100
PRO-ACTIVE
LABO 126
PRO-ACTIVE
LABO 288
PRO-ACTIVE/
ULTIMATE*
LABO 340
PRO-ACTIVE/
ULTIMATE*
Capacity in litres 95 95 280 330
Temperature setting
approx. in °C
approx. 2 to +15 approx. 2 to +15 approx. 0 to +15 approx. 0 to +15
Voltage in V 220–240 220–240 220–240 220–240
Frequency in Hz 50/60 50 50 50
Refrigerant quantity
in grams
43 32 30 40
Power consumption
in watts
30 105 88 110
Normal consumption
in kWh/24
0.33 0.58 0.68 0.65
Permissible ambient
temperature in °C
+10 to +38 +10 to +38 +10 to +38 +10 to +38
Specifies the interior
temperature
increase in minutes
at 25°C ambient tem-
perature in the event
of a power failure
86 44 55
Temperature homo-
geneity at 25°C ref-
erence ambient tem-
perature in K.
1.45
Temperature con-
stancy at reference
ambient temperature
of 25°C in K.
0.50
External dimensions
including wall dis-
tance (WxDxH) in cm
54 x 54 x 82 54 x 53.5 x 88.8 67 x 72 x 132 67 x 66 x 189
Usable dimensions
(WxDxH) in cm
44 x 43 x 34 43 x 33 x 59
(usable depth
below 15 cm less)
52.7 x 39 x 92
(usable depth
below 15 cm less)
52.7 x 39 x 116
External dimensions
with door open 90°
(WxD) in cm
54 x 106 54 x 106 67 x 130 67 x 124
Shelf size (WxD) in
cm
43 x 33 43 x 33 52 x 39 52.7 x 39
Appendix
Technical data
23.05.2023Refrigerator/freezer LABO series74