7. Turn unit key to position “1”.
ð
The test function is completed, the electronic delay is
switched on again.
The Display displays the current temperature of the
chilled goods.
The test function is ended automatically after
10 minutes.
8.
After the check, clean and disinfect the unit
Ä
Chapter 8
‘Cleaning and disinfection’ on page 56.
Remove the dust from the condenser (Fig. 6) at least every six
months in order to avoid affecting the performance of the cooling
machine.
Requirement:
n The back of the unit is accessible.
1. Dust off the condenser (for example with a brush or vacuum
cleaner).
2. Check the condenser for visible damage and wear.
Checking the condenser
Maintenance
Safety inspection
23.05.2023 Refrigerator/freezer LABO series 61