INSTALLATION INSTRUCTIONS Installation
10MAR06
II4825-1
Page
68 of 72
2 Touch [CREATE 35 x 43 cm SMPTE].
3 Send the SMPTE test pattern to a each PRINTER.
4 Locate a DENSITOMETER at the customer site.
5 From the PRINTER, use the DENSITOMETER to measure the density of the gray
patches of the SMPTE test pattern from 0% - 100%, .
6 Record each gray patch measurement.
7 Compare the measurement to the AIM
density and the tolerance.
8 Do all measured densities match the AIM density +/-0.10?
9 From the system, touch [Back].
SMPTE
Gray Patch
AIM
Density
Tolerance
0% 3.00 +/-0.10
10% 2.75 +/-0.10
20% 2.44 +/-0.10
30% 2.16 +/-0.10
40% 1.88 +/-0.10
50% 1.61 +/-0.10
60% 1.33 +/-0.10
70% 1.05 +/-0.10
80% 0.77 +/-0.10
90% 0.49 +/-0.10
100% 0.21 +/-0.10
Yes No
Continue with Step 9. a. Check that the MIM “Destinations” are set correctly.
b. Ask the customer to make a call to the equipment
service representative for the PRINTER.
c. Continue with Step 9.