INSTALLATION INSTRUCTIONS Installation
10MAR06
II4825-1
Page
70 of 72
12 Record each gray patch measurement.
13 Compare the measurement to the AIM
luminance and the tolerance.
14 Do all measured luminance match the AIM luminance +/-0.05?
15 From the “Diagnostic Menu”, run the [Autolooper Test] 10 times for each size
CASSETTE.
16 Use the Kodak DirectView TOTAL QUALITY TOOL to test and adjust the output image
quality.
17 Check that the software version is 3.0 or higher for the ROP. For more information, see
the INSTALLATION INSTRUCTIONS for the Kodak DirectView REMOTE OPERATIONS
PANEL (ROP) 5E2249.
18 Do the Field Service Installation Checklist for the CR SYSTEM 6F5825.
SMPTE
Gray Patch
AIM
Luminance
Tolerance
0% 0.12 +/-0.05
10% 0.35 +/-0.05
20% 0.76 +/-0.05
30% 1.46 +/-0.05
40% 2.71 +/-0.05
50% 4.86 +/-0.05
60% 8.59 +/-0.05
70% 15.26 +/-0.05
80% 26.36 +/-0.05
90% 46.36 +/-0.05
100% 80.00 +/-0.05
Yes No
Continue with Step 15. • Ask the customer to make a call to the equipment
service representive for the WORKSTATION to do a
calibration of the WORKSTATION.
• Continue with Step 15.