10 Operating Instructions DWM 2000 D
2.4 Parameters stored in the EEPROM (menu 2.4.3)
Paramete
Comment Typical value
CHECKS 1 Check EEPROM n°1 -
CHECKS 2 Check EEPROM n°2 -
CHECKS BYTE 1 Check EEPROM -
CHECKS BYTE 2 Check EEPROM -
CPT ALARM. Counting of all the error messages
since the first power-up.
-
CPT ALARM.2: Counting of all the error messages
since the last reset.
-
TEST Indication of auto diagnostic test
level
CUSTOM
CORRECTION
YES/NO
Indication of activation of low velocity
linearization.
YES
EP CHECKS Result of the data check in EPROM
(see menu 1.1.6.)
Value = 22309
With soft V1.02
FS Display of the programmed full scale
values (see menu 1.1.2.)
1 m/s ... 8 m/s
GK Primary head calibration constant
(see menu 2.1.3.)
0.8 ... 1.3
U REF Internal voltage reference
(see menu 1.1.3.)
2.4. ... 2.6.
T CST Time constant (see menu 2.2.4.) 5 s
TEST AMP Test value of the amplifier control
loop (see menu 1.1.4.)
75 ... 95
FM Frequency of the magnetic field
(see menu 1.1.1.)
10 ... 14.5 Hz
DIAMETER Diameter of pipeline
(see menu 2.4.6)
≥ 50 mm