INSTALLATION 2
7
LT40 C/F
www.krohne.com10/2021 - 4009051601 - AD UKEX LT40 R01 en
2.1 Precautions
2.1.1 General notes
You can download the certificate from our website.
2.1.2 Special conditions
Make sure that:
• you can get access to the device,
• there is sufficient space around the device for inspections
• you can see the device nameplate, and
• there are no external forces or shocks applied on the device.
WARNING!
When you install the device, obey the conditions in the UK-Type Examination certificate. These
conditions include:
•
The special conditions for safe use.
•
The Essential Safety Requirements.
DANGER!
This installation must agree with EN 60079-14: Explosive atmospheres – Part 14: Electrical
installations design, selection and erection.
WARNING!
EPL Ga equipment
The aluminium housing of the device is a possible source of ignition in a potentially explosive
atmo
sphere.
Install the device in an applicable location to make sure that iron/steel objects do not hit or rub
against aluminium parts.