The installations site for the unit must be draught-free and should be selected so that
frequent passing of people in front of the work opening is avoided.
1. Adjust the levelling screws to assure that the table top is in horizontal position and levelled.
2. Valves for gases or vacuum are installed in the side windows. A qualified technician must
make the connections for the supply.
When the cabinet has been installed:
1. Check that the front window is in the correct position, i.e. parallel to the front shield.
2. Never lift the front window manually - this will cause window to mall function.
3. Check that the window slides from top to bottom without any irregular sound.
4. If the window does not move when the UP or DOWN button is pressed, authorized service
personnel must be called to correct the fault.
5. Always keep hands and arms away from the working chamber when activating the front
window.
5.2 Connections
Required fusing: Circuit breaker 10 A (13 A) or fuse T 10 A (T 13 A).
In addition, the applicable safety requirements of the local power Supply Company shall apply.
If an automatic relay for disconnection of power in case of electric fault is needed, it
must be installed in the supply, as it is not built-in.
Before connecting the power supply it must be checked that the mains specifications
correspond to those stated on the type plate.
The safety cabinet is provided with a flexible power cord. The connection may be
installed hard-wired or by means of a wall outlet with protective ground.