Instruction manual for Fortuna
Item No. 700745
CopyRight © 2017
10 Quality testing of the Fortuna cabinets
All Fortuna cabinets are adjusted and tested before leaving the factory. A factory check list and a
factory test report is issued for each cabinet and is, together with the HEPA filter test certificates and
an instruction manual, shipped with the cabinet.
10.1 Test terms and definitions
Default value as specified by LaboGene.
Value measured at the location of the sterile cabinet.
Acceptable deviation from the nominal value.
The sum of the measuring values divided by the number of tests.
The average value is compared to the nominal value.
Acceptable operating value for the downflow air velocity.
Velocity of the displacement airflow circulating through the working
chamber.
10.2 What is required by EN 12469 to be tested
Leak test of HEPA filters
Downflow air velocity and setting of alarms
Special tests not included as default
Particle counting in working chamber
10.3 The factory quality test of the Fortuna cabinets
The factory quality test of the cabinets is designed to ensure that the cabinet is functioning in
accordance with requirements and specification before being shipped to the customer.
10.4 The factory check list and factory test report
The factory check list and factory test report is designed for the Fortuna cabinets specifically. They
document all quality related activities performed during the production through testing and cleaning
when the cabinet is ready for packing and shipment.
A filled in and reviewed check list and test report is shipped with each cabinet.
10.5 Test equipment needed to perform installation tests and service tests
The test equipment needed to perform installation and service test are: