In this case a PIR sensor can be mounted – let the cabinet run at full speed for defined time – as
long as there are movements. If no movements it will stop after the pre-set time from 60 seconds
and up.
In order to ensure the clean environment inside the cabinet, and that the cabinet instantly provides
operator safety – the cabinet has to run at reduced speed at all times during working hours
(programmable through the autostart – menu).
6.5 Setting of light
The light setting of the cabinet upon alarm can be set in 3 different alternatives:
Light always on when blue light button is activated.
Light always on when blue light button is activated and in full speed mode – if reduced
speed is pressed the light turns out.
Light is always on when blue light button is activated and in full speed mode – if there is an
alarm the light will turn off automatically and will only be turned on again when all alarms
have been deactivated.
6.6 Start-up of the Mars and Mars Pro cabinets
Turn on the cabinet by pressing the green button (1). The fans will start and the visible alarm in the
red button (3) will flash and audible alarm will be running until safe air velocity in downflow and inflow
has been reached. When the safe condition is reached the LED in the green button (1) will turn on.
The display will show the message below. Press the blue ENTER button to clear this message.
Turn on the light on in the cabinet by pressing the blue button (4). Move the window to working
position using the arrow UP (7) or DOWN (0) buttons.
The cabinet is ready for use with protection of product.
6.7 Operating modes of the Fortuna cabinets
The following are the operating modes for Fortuna cabinets:
OFF mode
Working mode
Standby mode
UV mode
6.7.1 Off mode
Off mode means the cabinet is in an idle state. The cabinet is turned off (light and fans are off). The
cabinet is connected to a power source.