EasyManuals Logo

LaboGene Fortuna 900 User Manual

LaboGene Fortuna 900
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #30 background imageLoading...
Page #30 background image
Instruction manual for Fortuna
12062017QA01
Item No. 700745
CopyRight © 2017
Page 30 of 49
Warning
As this cabinet can be used for processing and treating infectious substances, it may
become contaminated. Prior to disposal, the entire cabinet with filters must be
decontaminated in accordance with acceptable standards and procedures.
9 Specifications
Fortuna 900
Unit
Part no: 702904
General Specifications
Certificates
None
Compliance with
ISO 14644, EN 61010-1, EN 61000-6, EN 11201
Sound Pressure Level at factory settings and standard configuration
dB(A)
≤45
Sound Pressure Level in Eco-Save mode
dB(A)
≤42
Down flow, Air velocity
m/s
0,32 ±0,01
Light illuminance (fluorescent tube)
Lux
> 1000
Front Opening (height)
mm
350 (Optionally 550)
Table top
Flat, one piece
Filter construction in accordance with EN 1822-1
1-filter system, HEPA, at least Class H14
Table top and trough material
AISI 304 stainless steel
Table top max. total load
kg
100
Side glass material
5 mm hardened glass
Front glass material
6 mm laminated safety glass
Cabinet and back wall material/colour
Polyester coated steel/RAL 9016
Dimensions
External dimensions with adjustable Stand (Standard) (DxWxH)
mm
797 x 1004 x 2033
Table top height with adjustable Stand (Standard)
mm
750 (Range 745-780)
External dimensions with Elec. Elevation Stand (DxWxH)
mm
797 x 1004 X 2025-2325
Table top height with Elec. Elevation Stand
mm
750-950
Work chamber dimensions (DxWxH)
mm
650x900x700
Net weight
kg
157
Shipping Volume
m
3
Approx. 2,2
Electrical Data
Rated AC Voltage/frequency
V/Hz
220-240/50-60
Rated max. current (fuse)
A
Max. 10
Power outlet(s) max. total current (fuse)
A
6,3
Type of power outlet socket, pair left & pair right side
DK - type K (IP44)
Typical power consumption excl. power outlets
VA/W
57/52
Typical power consumption in Eco-Save mode
VA/W
30/21
Type of mains connection
3 pin male AC power socket

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LaboGene Fortuna 900 and is the answer not in the manual?

LaboGene Fortuna 900 Specifications

General IconGeneral
BrandLaboGene
ModelFortuna 900
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals