Instruction manual for Fortuna
12062017QA01
Item No. 700745
CopyRight © 2017
Page 17 of 49
By pressing 8 or 9 the user can move to Display mode where there is a possibility for
constant read out of velocity speeds both inflow and down flow showed in m/s or in
counts from 0-255.
By pressing 8 or 9 again the user can see UV-light (if this is installed) and choose the time
for how long the UV light should be activated.
By pressing 8 or 9 again the user can choose Auto start auto stop. Here it is possible to
select Monday to Friday whole week or only a single day when the cabinet should start
up in full speed and after how long it should wait until it switches off again.
By pressing 8 or 9 the Time setting on the cabinet can be set by the end user date,
month, and year, clock 24 hours or AM/PM.
By pressing 8 or 9 Password shows up this means the user can select a Password so
the keys cannot be activated without the entered Password. It can always be overruled
by the Service access code.
By pressing 8 or 9 Service mode can be reached. To enter this you need to be trained
personnel and to have got the Service access code.