EasyManua.ls Logo

LaboGene ScanSpeed 1248 - Positioning of Sample Tubes

Default Icon
21 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cat. No. 9.900.900.758
CopyRight © 2013
Page 12 of 20
If there is a water drop or dirt in the rotor hole or inner adaptor, remove it with soft dry cloth.
2. Tubes should be placed in the rotor with same amount of samples at symmetrical positions.
Only use appropriate centrifugal tubes and do not exceed the speed beyond the tubes max. g-force.
For safety, fill the sample for 70~80% in the tubes.
Correct Way of Sample
Balancing & Tube Usage
If the number of samples is not in pair,
please load the control tubes at each symmetrical
position. Otherwise, it results noise and vibration, which eventually damage the instrument.
For safety, the Imbalance Cut Off function will be occurred, if there is imbalance of loading tubes (Error 8,
Imbalance error). Please refer to 6. Trouble Shooting.

Related product manuals