Do not place dangerous materials within 30 cm distance around the instrument, and that is also
recommended by IEC 61010-2-020.
Use the emergency door open function only when the door button on the control panel is dumb under the
condition of complete stop of rotor running.
Never try to open or move the instrument if it is not completely stopped.
If the power input is more than ±10 % of the recommended voltage or fluctuates frequently, it may cause
malfunction of the instrument and often result serious damage.
Install the instrument at the place without any kinds of corrosive gases.