SPECIFICATIONS
Instrument Type Nephelometric turbidity, calibrated in NTU
Range: 0.00 -1100 NTU
Accuracy: .05 or ±2% for readings below 100 NTU, whichever is
greater
±3% above 100 NTU
Resolution: Standard Mode
0.01 from 0.00 -10.99 NTU
0.1 from 11.0 -109.9 NTU
1 from 110 -1100 NTU
EPA Mode
NTU Reported to the nearest NTU
0 - 1.0 0.05
1.0 - 10 0.1
10 - 40 1
40 - 100 5
100 - 400 10
400 - 1000 50
1000 100
Display: 3½ digits
Response Time: 5 seconds
Warm-up time: Not required
Automatic Shut Off: 2 minutes
Lamp: Tungsten Filament bulb (approximate life 800 hours)
Sample: 15 mL in capped tube
Sample Chamber: Accepts 25mm diameter flat-bottomed tubes (capped)
Power source: Battery Operation: 9 Volt Alkaline
Line Operation: 120V/50Hz, 220V/60Hz*, with supplied
adapter
Size (L X W X H): 8.5 x 16.2 x 6.7 cm, 3.4 X 6.4 X 2.6 inches
Shipping Weight: Meter only: 11 oz. (312g)
Kit: 3 lb. 7 oz. (1560g)
Serial Interface: RS232, 8 pin mDIN, 9600b, 8, 1, n
*CE Mark: The devise complies to the product specifications for the Low Voltage Directive
when furnished with the 220V AC Adapter (Code 1774). The 120V AC adapter
is not CE approved.