- 67 -
SWITCHING FROM GROSS TO NET WEIGHT
The PC transmits the ASCII string: $aaNETckckCR
Possible instrument responses:
- correct reception: &&aa!\ckckCR
- incorrect reception: &&aa?\ckckCR
SWITCHING FROM NET TO GROSS WEIGHT
The PC transmits the ASCII string: $aaGROSSckckCR
Possible instrument responses:
- correct reception: &&aa!\ckckCR
- incorrect reception: &&aa?\ckckCR
READING OF DECIMALS AND DIVISION NUMBER
The PC transmits the ASCII string: $aaDckckCR
Possible instrument responses:
- correct reception: &aaxy\ckckCR
- incorrect reception: &&aa?\ckckCR
where: x ................... number of decimals
y = 3 ............. for division value = 1
y = 4 ............. for division value = 2
y = 5 ............. for division value = 5
y = 6 ............. for division value = 10
y = 7 ............. for division value = 20
y = 8 ............. for division value = 50
y = 9 ............. for division value = 100
TARE ZERO-SETTING
The PC transmits the ASCII string: $aazckckCR
where: z ................... command of weight zero-setting (122 ASCII)
Possible instrument responses:
- correct reception: &aaxxxxxxt\ckckCR
- incorrect reception: &&aa?\ckckCR
- the gross weight is not displayed on the instrument: &aa#CR
where: xxxxxx ...... 6 characters to indicate the required weight value
t ................... character to indicate the weight (116 ASCII)