5-2 Performance Verification WR6K Rev. D
5.2 Test Equipment Required
These procedures use external, traceable signal generators, DC precision power
supply and digital multi-meter, to directly check specifications.
Instrument Specifications Recommended
Signal Generator
Radio Frequency
Frequency : .5 MHz to 3 GHz
Frequency Accuracy : 1 PPM
HP8648B/C, Fluke 9500 or
equivalent
Signal Generator
Audio Frequency
Frequency : 0 to 5 kHz
Amplitude : 8 V peak to peak
HP33120A or equivalent
Voltage Generator
DC Power Supply
Range of 0 to 20 V, in
steps of no more than 15 mV
HP6633A or equivalent
Power Meter +
Sensor
Accuracy ±1 %
HP437B + 8482A or
equivalent
Digital Multimeter
Volt & Ohm
Voltmeter Accuracy : 0.1 %
Ohmmeter Accuracy : 0.1 %
Keithley 2000 or equivalent
Coaxial Cable, 5 ns
50Ω, BNC, length 100 cm,
Coaxial Cable, 5 ns
50Ω, SMA, length 100 cm,
2 Attenuators, 20 dB
50Ω, BNC, 1 % accuracy
T adapter
50Ω, BNC T adapter
Table 5-1 : Test Equipment
5.2.1 Test Records
The last pages of this document contain WaveRunner 6000 series test records in the
format tables. Keep them as masters and use a photocopy for each calibration.
5.3 Turn On
If you are not familiar with operating the WaveRunner 6K series, refer to the
operator's manual.
Switch on the power using the power switch.
Wait for about 20 minutes for the scope to reach a stable operating
temperature:
Insert the CD with panel setups