Trigger on : EXT
Impedance : DC 1M
Connect the DMM to External, and measure the input impedance, reverse the
meter leads and measure the input impedance.
Average these two numbers and record the input impedance in Table 2, and
compare the result to the limit in the test record.
5.5 Leakage Current
Specifications
DC 50, EXT DC 50 : ±2 mV
The leakage current is tested by measuring the voltage across the input channel.
5.5.1 Channel Leakage Current
a. DC 50
Recall Leakage - 50 ohm x1.lss or configure the DSO :
Panel Setups : Recall FROM DEFAULT SETUP
Channels Trace ON Channel 1, Channel 2, Channel 3 & Channel 4
Input Coupling : DC 50 on all 4 Channels
Input gain : 20 mV/div. on all 4 Channels
Trigger mode : Auto
Time base : 50 nsec/div.
WR6K Rev.D Performance Verification 5-7