5-8 Performance Verification WR6K Rev. D
Set the DMM to measure Volts, and connect it to Channel 1.
Measure the voltage and enter it in Table 3. Compare it to the limits.
Repeat the test for all input channels.
Recall Leakage - 50 ohm x10.lss or set Input gain to 200 mV/div. on all 4
Channels
Repeat the test for all input channels. Record the measurements in Table 3,
and compare the results to the limits in the test record.
b. DC 1M
Recall Leakage - 1 Mohm x1.lss or configure the DSO :
Panel Setups : Recall FROM DEFAULT SETUP
Channels Trace ON Channel 1, Channel 2, Channel 3 & Channel 4
Input Coupling :
DC 1M on all 4 Channels
Input gain : 20 mV/div. on all 4 Channels
Trigger mode : Auto
Time base : 50 nsec/div.