Set the DMM to measure Volts, and connect it to Channel 1.
Measure the voltage and enter it in Table 3. Compare it to the limits.
Repeat the test for all input channels.
Recall Leakage - 1 Mohm x10.lss or set Input gain to 200 mV/div. on all 4
Channels
Repeat the test for all input channels. Record the measurements in Table 3,
and compare the results to the limits in the test record.
Recall Leakage - 1 Mohm x100.lss or set Input gain to 2 V/div. on all 4
Channels
Repeat the test for all input channels. Record the measurements in Table 3,
and compare the results to the limits in the test record.
WR6K Rev.D Performance Verification 5-9