Check that the Sampling rate is 10 GS/s
Press Clear Sweeps.
Measure for at least 50 sweeps, then press Stop to halt the acquisition.
Record the two max pkpk of Ch1 & Ch3 in Table 4, and compare the test results
to the limits in the test record.
b) Channel 2 & Channel 4
Recall Noise 10GS Ch2 & Ch4.lss or configure the DSO as shown in 5.6.1.a.
and make the following changes :
Input Coupling : DC 50 on all 4 Channels
Input gain : 10 mV/div. On all 4 Channels
Channels Trace ON Channel 2, Channel 4
Channels Trace OFF Channel 1, Channel 3
Time base : 500 ηsec/div.
Select Time base Setup
Channel use : Automatic
Press : Measure
Change parameters
P1 : Turn Off
P2 : Measure pkpk of Ch4
P3 : Turn Off
P4 : Measure pkpk of Ch4
WR6K Rev.D Performance Verification 5-15