5-14 Performance Verification WR6K Rev. D
5.6.2. 10 GS/s Channel Mode (WR6100 & WR6200 only)
a) Channel 1 & Channel 3
Recall Noise 10GS Ch1 & Ch3.lss or configure the DSO as shown in 5.6.1 and
make the following changes :
Input Coupling : DC 50 on all 4 Channels
Input gain : 10 mV/div. On all 4 Channels
Channels Trace ON Channel 1, Channel 3
Channels Trace OFF Channel 2, Channel 4
Time base : 500 ηsec/div.
Select Time base Setup
Channel use : Automatic
Press : Measure
Change parameters
P1 : Measure pkpk of Ch1
P2 : Turn Off
P3 : Measure pkpk of Ch3
P4 : Turn Off