Press Clear Sweeps.
Measure for at least 50 sweeps, and then press Stop to halt the acquisition.
Record the four max pk-pk parameter values in Table 4, and compare the test
results to the limits in the test record.
Repeat the test for Time base : 1 msec/div.
Record the measurements (high pk-pk of 1,2,3,4) in Table 4, and compare the
results to the limits in the test record.
WR6K Rev.D Performance Verification 5-13