5-12 Performance Verification WR6K Rev. D
5.6 Peak-Peak Noise Level
Description
Noise tests with open inputs are executed on all channels with 0 mV offset, at a gain
setting of 10 mV/div. The scope parameters functions are used to measure the Peak to
Peak amplitude of the noise.
Specifications
7.5 mV Peak-Peak at 10 mV/div.
5.6.1 5 GS/s 50 Ohm
With no signal connected to the inputs
Recall Noise - 10mv.lss or configure the DSO :
Panel Setups : Recall FROM DEFAULT SETUP
Channels Trace ON Channel 1, Channel 2, Channel 3 & Channel 4
Input Coupling : DC 50Ω on all 4 Channels
Input gain : 10 mV/div. on all 4 Channels
Input offset : 0.0 mV on all 4 Channels
Trigger setup : Edge
Trigger on : C1
Trigger Mode : Auto
Time base : 1 µsec/div.
Channel use : Automatic
Press : Measure, Measure Setup,
Statistics : On
Add parameters
P1 : Measure pkpk of Ch1
P2 : Measure pkpk of Ch2
P3 : Measure pkpk of Ch3
P4 : Measure pkpk of Ch4