b. DC 1M
Recall Leakage - 1 Mohm ext x1.lss or configure the DSO as shown in 5.5.1
and make the following changes:
Select Setup trigger
Set Trigger on : EXT
Impedance : DC 1M
Connect the DMM to External.
Measure the voltage and enter it in Table 3. Compare it to the limits.
WR6K Rev.D Performance Verification 5-11