EasyManua.ls Logo

LeCroy WAVERUNNER 6000 SERIES - Page 76

Default Icon
325 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-34 Performance Verification WR6K Rev. D
Increase the frequency of the generator to the maximum input frequency for that
model, adjusting the amplitude so that the power remains constant and measure
the value of Sdev. Record in Table 13.
Repeat the above 3 steps for Channel 2, (Bandwidth – CH2 10mv.lss) 3
(Bandwidth – CH3 10mv.lss) & 4 (Bandwidth – CH4 10mv.lss) substituting
channel controls and input connector. Record the measurements in Table 13.

Related product manuals