5-36 Performance Verification WR6K Rev. D
5.10 Trigger Level
5.10.1 Description
The trigger capabilities are tested for several cases of the standard edge trigger:
Channel (internal), and External Trigger sources
Three DC levels: 2.5, 0, +2.5 major screen divisions
Positive and negative slopes
5.10.2 Channel Trigger at 0 Division Threshold
Recall Trigger - CH1 0 div neg slope.lss or configure the DSO:
Panel Setups : Recall FROM DEFAULT SETUP
Channels Trace ON Channel 1, Channel 2, Channel 3 & Channel 4
Input Coupling : DC 50 on all 4 Channels
Input gain : 100 mV/div. on all 4 Channels
Input offset : 0 mV on all 4 Channels (use show status to verify)
Trigger setup : Edge
Trigger on : C1
Slope 1 : Pos
Mode : Auto
Set Trigger level : DC 0.0 mV
Pre-Trigger Delay : 50 %
Time base : 1 µsec/div.
C1 Pre-Processing: Averaging 10 sweeps
C2 Pre-Processing: Averaging 10 sweeps
C3 Pre-Processing: Averaging 10 sweeps
C4 Pre-Processing: Averaging 10 sweeps
Set the output of the sine wave generator to 100 KHz.
Connect the output of the generator to Channel 1 through a 50 Ohm coaxial cable
as shown in Figure 5-10 and adjust the sine wave output amplitude to get 90% of
full scale.