Rev. D Performance Verification   5-29
!
Set the output of the LeCroy LW420 or equivalent audio frequency signal 
generator to 
1 kHz.
!
Connect the output of the generator to Channel 1 through a 50 Ohm coaxial 
cable and adjust the sine wave output amplitude
 
to get
 8 divisions peak to 
peak .
!
Select WAVE PILOT :
Cursors, Time, Absolute
!
Use the "cursor position" knob, to move the 
Time marker
 at 0.0 
µ
s
 
!
Press 
Clear Sweeps
,
!
Acquire 10 sweeps and record in Table 10 the 
level
 readout displayed below 
100 mV in the icon
 1, 
at top left
.
!
Compare the test results to the corresponding limit in the test record.
!
  Set Trigger Slope 1 : 
Neg
!
  Acquire 10 sweeps and record in Table 10 the 
level
 readout displayed below
100 mV in the icon
 1, 
at top left
.