Rev. D Performance Verification   5-43
5.11 Smart Trigger
5.11.1 Trigger on Pulse Width 10 nsec
a. Pulse Width < 10 nsec
!
Recall 
xxxP046.PNL
  or configure the DSO
Panel Setups :
Recall FROM DEFAULT SETUP
Channels trace ON
Channel 1
Input coupling :
DC 50
Ω
ΩΩ
Ω
Input gain  : 
.5 V/div.
Input offset : 
0 mV
Trigger setup :
Smart
Setup Smart Trigger
Glitch
Trigger on :
1
At the end of :
Neg.
Width < 10 nsec :
On
Width > 10 nsec :
Off
Trigger mode :
Norm
Time base :
5 nsec/div.
!
Connect the 
RF output
 of the HP8648B generator through a 5ns 50 Ohm BNC
coaxial cable into Channel 1.
!
Set the generator frequency to
 100 MHz
. Adjust the generator output amplitude
to get
 5 divisions peak to peak .
!
Check that the scope Triggers. Record the test result in Table 12.
!
Set Width < 10 nsec  
Off 
and 
 
Width > 10 nsec 
ON
!
Check that the scope 
doesn’t trigger : 
slow trigger and no flashes in box next
to  normal. Record the test result in Table 12.
b. Pulse Width  > 10 nsec
!
Set the generator frequency to
 40 MHz
.
!
Set Width < 10 nsec  
Off 
and 
 
Width > 10 nsec 
ON
!
Check that the scope Triggers. Record the test result in Table 12.
!
Set Width < 10 nsec  
On 
and 
 
Width > 10 nsec 
Off
!
Check that the scope 
doesn’t trigger : 
slow trigger and no flashes in box.