EasyManua.ls Logo

Leica RM2145 - 10. Cleaning

Leica RM2145
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
Instruction manual V3.2 – 02/2002
10. Cleaning
Before cleaning, lock the handwheel, switch
the instrument off and pull the mains plug!
Only use mild household cleaners / soap for
cleaning.
The varnish paint coat of the instrument is not
resistant to acetone or xylene!
Make sure that no liquids get into the interior
of the instrument while cleaning!
Remove sectioning debris with a dry brush.
Pull out the section waste tray to empty.
Remove the arm or hand rests (optional accesso-
ries) for cleaning.
Remove knife holder and knife holder base.
Clean the instrument with a damp (not wet!) cloth.
After cleaning wait for the instrument to be
completely dry before switching it back on!