Troubleshooting 8-3
D-H021.1011EN
innovators in agriculture
Code Description Result Reason Action
The lid of the bottle to
be filled is not fully open
1. Push the button
[STOP/RESET] for a
short time.
2. Make sure the lid is
between the racks
3. Push and hold
button [STOP/RESET]
for more than 5 s.
The bottle to be filled is
not installed correctly
1. Push the button
[STOP/RESET] for a
short time.
2. Make sure all bottles
are installed correctly
3. Push and hold
button [STOP/RESET]
for more than 5 s.
The rack is not straight
or bent
1. Push the button
[STOP/RESET] for a
short time.
2. Make sure the rack is
undamaged.
3. Push and hold
button [STOP/RESET]
for more than 5 s.
The cover of the
SHUTTLE is not closed
correctly
1. Push the button
[STOP/RESET] for a
short time.
2. Close the cover of
the SHUTTLE
3. Push and hold
button [STOP/RESET]
for more than 5 s.
5
The fill head could
not move down
(Z-axis) or could
not reach the
lowest position
three times
The sampling
process stops
The racks are not
positioned correctly
Make sure the racks are
positioned correctly
The rotation sensor
(encoder) is broken
1. Push the button
[STOP/RESET] for a
short time.
2. Turn the spindle
manually, very slowly
(approx. 1 revolution
in 2 s)
3. Push and hold
button [STOP/RESET]
for more than 5 s.
4. Make sure the fill
head moves to the
home position.
6 The controller
does not get
feedback from the
X-axis motor
The sampling
process stops
The fill unit is stuck
1. Push the button
[STOP/RESET] for a
short time.
2. Turn the spindle
manually, very slowly