LD5208 USER MANUAL, REV. 0: NOVEMBER 2005
20
the needs of each application.
Data output is effected via COM1 (RS232C) and can be driven to a printer (TM-295 or Seikosha SP-
2400 compatible) or to a PC.
Data output format is selected via SETUP 2, parameter 2.t or functions 21-30 (see paragraph 9.2)
Errors that may occur during the printing are listed in chapter 13.
10.5.1 To print a weigh ticket :
1. Wait for the scale to stabilize (respective status annunciator ON)
2. Press the Print key )
3. The display shows P r I n t briefly and returns to the normal weight display
10.5.2 To produce a copy of the last weigh ticket :
If the first weigh ticket has not been printed clearly, or the paper has been damaged, you can
produce a copy of the last weigh ticket,
.
..
. provided no other action has been taken after the last print (no other key has been
pressed)
1. Press key > and keep pressed for about 3sec
2. Display shows F n 0 0 . Type in the number 43, using key ~ to increment the flashing digit
and key > to move one digit to the right
3. Press key ( to accept the value
4. The display shows P r I n t briefly and a copy of the last weigh ticket is printed
.
..
. The last printout is lost in power down
10.5.3 To temporarily disable the printer output :
If for some reason you wish disable the print function, then
1. Press key > and keep pressed for about 3sec
2. Display shows F n 0 0 . Type in the number 20, using key ~ to increment the flashing digit
and key > to move one digit to the right
3. Press key ( to accept the value
.
..
. Printer output is enabled automatically on power up