Lexicon
5-23
<S:0000F R:FFFF0
LABUS1 shorted to ground- Add Test A:00000
<S:0000F R:0202F
LABUS2 shorted to ground- Add Test A:00000
<S:0000F R:0404F
LABUS3 shorted to ground- Add Test A:00000
<S:0000F R:0808F
LABUS4 shorted to ground- Add Test A:00000
<S:0000F R:1010E
LABUS5 shorted to ground- Add Test A:00000
<S:0000F R:2020D
LABUS6 shorted to ground- Add Test A:00000
<S:0000F R:4040B
LABUS7 shorted to ground- Add Test A:00000
<S:0000F R:80807
LABUS8 shorted to ground- Hi Addr A:40000
<S:40404 R:10101
To exit, press the OPTIONS button and the previous selection will be displayed.
WCS Failure (12)
When the WCS Test fails, Address, Data Sent and Data Received fields are used. All of these fields will
display the information in hex. The Address field will display what the Lexichip Address (LA0-LA9) was at
the time of the failure. The Data Sent field will display what Data was sent to the Writeable Control Store
(program memory) of the Lexichip. The Data Received will display the Data received from the WCS.
When Test Result is selected, the display will read the following:
Last Test: Failed
<> *1 times
This assumes that the test was run 1 time.
Pressing the >YES button 2 times will display the following if the WCS Test failed:
(This is an example where LA3 is shorted to ground.)
Test #12 A:5000
<S:5000 R:5008