6
1. Setup
Standard Items Included
Before setting up your new EVOS
XL Core Imaging System, unpack the unit and
accessories, and verify all parts are present. Contact your distributor if anything is
missing.
Note: If you do not have your distributor information, contact Technical
Support (see page 36).
• EVOS
®
XL Core Imaging System (includes fixed
stage)*
• Wireless mouse
• Power adaptor
• USB flash drive (includes User Guide and Quick
Start Guide)
• Condenser shield, removable
• Glass stage plate (packed in bubble wrap)
• Dust cover
* The attachable Mechanical Stage is an optional feature
available separately from Life Technologies. For
ordering information, contact your distributor or
Technical Support (see page 36).
Condenser shield,
removable
Operating Environment
• Place the instrument on a level surface away from vibrations from other pieces
of equipment.
• Allow at least 5 cm (2 in) free space at the back of the LCD housing to allow
for proper ventilation and prevent overheating of electronic components.
• Set up the EVOS
®
XL Core Imaging System away from direct light sources,
such as windows. Ambient room lighting can enter the imaging path and
affect the image.
• Operating temperature range: 4°C–32°C (40°C–90°F).
•
Relative humidity range: 30–90%.
IMPORTANT
! EVOS
®
XL Core Imaging System should not be subjected
to UV sterilization. UV degrades many materials, including plastic.
Damage from UV exposure is not covered under the manufacturer’s