14
Lifeloc FC10
Plus
www.Lifeloc.com | 303-431-9500
Setting Trigger Mode
From the “SETTINGS” display, press the Execute button.
Press the Function button until the display reads “TRIGGER MODE”.
Press the Execute button to select between “END OF BREATH” and “PRECISE VOLUME”.
Press the Function button to save the setting. Momentarily press Power to return to the
testing mode.
(User settings fl ow chart is shown on p. 32)
Security Settings Explained
Setting Up Security Password
Password can be set to protect access to calibration, calibration check, calibration
settings and user settings.
Without a password, users are still able to conduct Automatic, Manual and Passive tests;
print test results; and check battery status and temperature.
Press the Function button until the display reads “SETTINGS”.
Press the Execute button. Display reads “TIME”.
Press the Function button until the display reads “SECURITY SETTINGS”.
Press the Execute button. The unit is now ready to accept a password. Use the + or -
buttons to enter a number or letter.
Use the Execute button to move to the next digit. Password is limited to 8 characters.
Press the Function button to save your password. Momentarily press Power to return to
the testing mode.
Note: Record and store your password in a safe place. Lifeloc does not have access to
your password. If password is lost, you will need to contact Lifeloc to restore a default
via remote diagnostics or by returning the unit to the factory.