Setting Results Format
Press the Function button until the display reads “SETTINGS”.
Press the Execute button. Display reads “TIME”.
Press the Function button until the display reads “DISPLAY SETTINGS”.
Press the Execute button. Display reads “LCD CONTRAST”.
Press the Function button until the display reads “RESULTS FORMAT”.
Press Execute to toggle between “NUMERIC” and “PWF”.
Press the Function button to save your setting.
Momentarily press Power to return to the testing mode.
Results Format Explained
Results for Auto and Manual tests can be displayed in a “NUMERIC” or “PASS/WARN/
FAIL” (PWF) format. Passive test results, however, can only be displayed as “POS” &
“NEG”.
NUMERIC results are in a 3-digit BAC format.
PASS/WARN/FAIL results require setting specifi c levels for Pass and Fail.
PASS
FAIL
For example, if Pass is set at .040 and Fail is set at .080
(default settings on your FC10
Plus
), then:
Pass = .000 - .040
Warn = .041 - .079
Fail = .080 and above
Sample settings only. See
page 18 for instructions on
setting the Pass/Warn/Fail
levels.
User Settings