On-screen Messages
Messages Explained
Message Explanation
<1.3L Retest or Try Manual Test Breath fl ow ended before the subject blew
1.3 liters of breath. Instruct subject to try
again or use manual test mode.
>0.6 BAC is unusually high and above 0.6 BAC.
Subject may require medical attention.
Calibration/Cal Check Expired Calibrate or Cal Check the unit.
Calibration/Cal Check Expiring The unit is within 48 hours of the
Calibration or Cal Check lockout time.
External Interference External interference has been detected.
Move to a different location and try again.
Flow Error - Retry & Blow Steadily Exhalation not complete or interrupted.
Instruct subject to blow steadily as long as
they can.
Invalid Calibration/Cal Check No alcohol was detected. Repeat setup,
then retest.
Log Empty There are no results in the memory.
Low Battery Battery voltage is too low to take a test.
Replace batteries.
Low Li Battery The internal clock battery is low. Contact
Lifeloc Technical Support.
Pump Reset Needed The pump needs to be reset. Follow
onscreen instructions.
Temperature When calibrating or taking a test, unit is
outside of temperature limits.
Calibration range = 68
o
- 95
o
F (20
o
- 35
o
C)
Testing range = 32
o
- 130
o
F (0
o
- 55
o
C)
Timeout User blew for > 15 seconds.
Try again, blow harder.