Specifications
Size Range 0.3 - 25.0 µm
Channel Thresholds Standard: 0.3, 0.5, 1.0, 5.0 µm
Optional: 0.3, 0.7, 1.0, 3.0, 5.0, 10.0 µm
Other sizes available; specify at time of order
Flow Rate 1.0 CFM (28.3 LPM)
Counting Efficiency 50% (per JIS)
Light Source Elite™ Laser Diode
Laser MTBF > 20 Years
Zero Count Level <1 count/5 minutes (per JIS)
Vacuum Pump Internal Vacuum Pump
Calibration NIST Traceable, per ISO 21501-4
Communication Modes Ethernet TCP/IP; RS-485 Modbus; RS-232
via USB
Data Storage 3000 records, rotating buffer
Supporting Software Lighthouse Monitoring System, LMS
XChange, LMS Express, LMS Express RT/RT
Plus
Analog Ports Optional: two ports; can handle up to four 4-
20mA sensors (see below)
Environmental Sensors Optional: Temperature / Relative Humidity,
Air Velocity, Differential Pressure
Power Input 100-240 VAC, 50-60Hz, 130W
Display Remote Display; color LCD Touchscreen
(optional)
Enclosure Stainless Steel
Dimensions 9.5"[l] x 6.5"(w) x 8.5"(h) [24.13 x 16.51 x
21.59 cm]
Weight 12.2 lbs (5.53 kg)
Operating Temp/RH 50° F to 104° F (10° C to 40° C) / 20% to 95%
non-condensing
Storage Temp/RH 14° F to 122° F (-10° C to 50° C) / Up to 98%
non-condensing
Table 2-1 REMOTE 3104P Specifications