EasyManua.ls Logo

Lighthouse REMOTE 3104P - Addressing (DIP 1-6)

Default Icon
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming
248083336-1 Rev 3 4-3
Addressing (DIP 1-6)
Note:
Because Address
’0’ is reserved for RS-485
broadcast. Whenever all
of the dip switches are
OFF or when DIP switch1
is ON, the instrument
address is set to ’1’.
Table 4-2 details the addresses set by the DIP switches 1-6.
Table 4-2 DIP Switch Addressing
DIP SWITCHES
1 2 3 4 5 6
ADDRESS
DIP SWITCHES
1 2 3 4 5 6
ADDRESS
0 0 0 0 0 0 1 0 0 0 0 0 1 32
1 0 0 0 0 0 1 1 0 0 0 0 1 33
0 1 0 0 0 0 2 0 1 0 0 0 1 34
1 1 0 0 0 0 3 1 1 0 0 0 1 35
0 0 1 0 0 0 4 0 0 1 0 0 1 36
1 01 0 0 0 5 1 0 1 0 0 1 37
0 1 1 0 0 0 6 0 1 1 0 0 1 38
1 1 1 0 0 0 7 1 1 1 0 0 1 39
0 0 0 1 0 0 8 0 0 0 1 0 1 40
1 0 0 1 0 0 9 1 0 0 1 0 1 41
0 1 0 1 0 0 10 0 1 0 1 0 1 42
1 1 0 1 0 0 11 1 1 0 1 0 1 43
0 0 1 1 0 0 12 0 0 1 1 0 1 44
1 0 1 1 0 0 13 1 0 1 1 0 1 45
0 1 1 1 0 0 14 0 1 1 1 0 1 46
1 1 1 1 0 0 15 1 1 1 1 0 1 47
0 0 0 0 1 0 16 0 0 0 0 1 1 48
1 0 0 0 1 0 17 1 0 0 0 1 1 49
0 1 0 0 1 0 18 0 1 0 0 1 1 50
1 1 0 0 1 0 19 1 1 0 0 1 1 51
0 0 1 0 1 0 20 0 0 1 0 1 1 52
1 0 1 0 1 0 21 1 0 1 0 1 1 53
0 1 1 0 1 0 22 0 1 1 0 1 1 54
1 1 1 0 1 0 23 1 1 1 0 1 1 55