Lighthouse REMOTE 3104P, 5104P Operating Manual
4-4 248083336-1 Rev 3
Communicating
with the
Instrument
The RJ-45 connector on the instrument marked "RS-485" in Figure 4-3
may be used to connect the unit to a desktop or laptop PC COM port
when the distance to the instrument is greater than 50 feet. It is,
however, specifically designed to connect the REMOTE 3104P or
5104P to a MODBUS RS-485 network. Whenever possible, it is
prudent to use the USB port for connection to a PC.
WARNING:
Only one
port may be used at a
time. The instrument
must be powered OFF
before
changing from
USB, Ethernet or RS-485
to any other port.
Figure 4-3 The Communication Ports
0 0 0 1 1 0 24 0 0 0 1 1 1 56
1 0 0 1 1 0 25 1 0 0 1 1 1 57
0 1 0 1 1 0 26 0 1 0 1 1 1 58
1 1 0 1 1 0 27 1 1 0 1 1 1 59
0 0 1 1 1 0 28 0 0 1 1 1 1 60
1 0 1 1 1 0 29 1 0 1 1 1 1 61
0 1 1 1 1 0 30 0 1 1 1 1 1 62
1 1 1 1 1 0 31 1 1 1 1 1 1 63
Table 4-2 DIP Switch Addressing
DIP SWITCHES
1 2 3 4 5 6
ADDRESS
DIP SWITCHES
1 2 3 4 5 6
ADDRESS